Synergy Nano Technical Manual, Revision -
Page 19
2.3.1 Process Inputs
The Synergy Nano has multiple process inputs. These are listed in the table below.
Synergy Nano
Inputs Channels Application
T/C, RTD channels 2 Temperature Measurements
High Resolution
0-5VDC channels, 16-Bit
2 Humidity, temperature and other process variables. 0.0001
Volt resolution
UUT Temperature Inputs
T-Type Thermocouples
64 max Temperature Measurements
Synergy Nano Plus
Inputs Channels Application
RTD channels 2 Temperature Measurements
High Resolution
0-5VDC channels, 16-Bit
4 Humidity, temperature and other process
variables. 0.0001 Volt resolution
Low Resolution
0-5VDC channels, 10-Bit
8 For refrigeration pressure and process variables.
0.005 Volt resolution
UUT Temperature Inputs
T-Type Thermocouples
16 per Module
64 max
Temperature Measurements
Note: All of these inputs can be logged and checked using the user programmable alarm system.
2.3.2 Input Calibration and Scaling
Each Input can be calibrated for the physical measurement, Ohms, Volts or Degrees C.
Each Input can also be scaled to appropriate engineering units. (See the Calibration section)
2.3.3 Channel Calibration
Each Channel is assigned an input. In addition, channel data can be scaled and offset to accommodate
Channel errors caused by sensor position, thermal gradients, etc.
2.3.1
Process
Inputs
2.3.2
Input Cal.and
Scaling
2.3.3
Channel
Calibration
2.3.4
Channel
PIDs
2.3.5
Channel
Primitives